Common technical data
Sheathing material:
Inconel 600 (2.4816)
Operational temperature:
- 200°C  up to + 1150 °C
Insulation:
Magnesium oxide, highly compressed
Wire material:
NiCr - Ni (Type K) to DIN IEC 584
Limit deviation:
better than category 1 ( < 1.5 ° C bzw. 0.4 % x T ), the higher value applies
Measuring point:
welded insulated from the sheath (TI)
Bending radius:
> 3-fold sheath diameter
Response time:
< 150 ms with 1.0 mm sheath diameter

Different diameters, lengths and types (e.g..: J, N, S, L, B, T, E) on request.


Structure of a Sheating thermocouple:




Code letter
EN 60584
IEC 584-3 mod.
DIN 43722
DIN 43710**
DIN 43714***
ANSI MC96.1
J*

K

te-gelb-gelb-rot
T* te-blau-blau-rot
E   te-pink-pink-rot
S/R
te-gruen-schwarz-rot
N
  te-orange-orange-rot
B
  te-grau-grau-rot

* according Type L and U reffered to DIN 43710.

** DIN 43710 retired since 07/1997.
*** DIN 43714 retired since 03/2000.



Electronic Sensor offer the following products: